Item type |
紀要論文 / Departmental Bulletin Paper(1) |
公開日 |
2019-04-03 |
タイトル |
|
|
タイトル |
Simulation using patient-specific hollow three-dimensional models for percutaneous closure of patent ductus arteriosus in adult: Complete republication |
|
言語 |
en |
その他(別言語等)のタイトル |
|
|
その他のタイトル |
患者データから自施設で作製した中空立体模型を用いた,成人動脈管開存の経皮的閉鎖術シミュレーション |
|
言語 |
ja |
言語 |
|
|
言語 |
eng |
キーワード |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
simulation |
キーワード |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
three-dimensional printing |
キーワード |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
patent ductus arteriosus |
キーワード |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
percutaneous closure |
キーワード |
|
|
言語 |
en |
|
主題Scheme |
Other |
|
主題 |
adult |
資源タイプ |
|
|
資源タイプ識別子 |
http://purl.org/coar/resource_type/c_6501 |
|
資源タイプ |
departmental bulletin paper |
著者 |
Kataoka, Koichi
Kawada, Masaaki
Matsubara, Daisuke
Oka, Kensuke
Suzuki, Shun
Minami, Takaomi
Imai, Yasushi
|
著者別名 |
|
|
|
姓名 |
片岡, 功一 |
|
|
言語 |
ja |
著者別名 |
|
|
|
姓名 |
河田, 政明 |
|
|
言語 |
ja |
著者別名 |
|
|
|
姓名 |
松原, 大輔 |
|
|
言語 |
ja |
著者別名 |
|
|
|
姓名 |
岡, 健介 |
|
|
言語 |
ja |
著者別名 |
|
|
|
姓名 |
鈴木, 峻 |
|
|
言語 |
ja |
著者別名 |
|
|
|
姓名 |
南, 孝臣 |
|
|
言語 |
ja |
著者別名 |
|
|
|
姓名 |
今井, 靖 |
|
|
言語 |
ja |
抄録 |
|
|
内容記述タイプ |
Abstract |
|
内容記述 |
Background and Objective:It is challenging to accurately predict the shape and stability of occlusion devices during percutaneous closure of patent ductus arteriosus (PDA) in adults. In this study, we evaluate the utility of percutaneous closure simulation of PDA in adults using patient-specific hollow threedimensional(3D) models created from computed tomography data. Methods:The study was conducted on four adult patients with PDA, whose age ranged from 54 to 75 years. While three were type E as per the Krichenko classification, one was type D. We created a transparent silicone hollow model using a solid model, which was printed by a personal 3D printer from contrast-enhanced multi-detector computed tomography data, as a mold. All processes were carried out at our institution. Before actually performing percutaneous closure of PDA, we simulated procedures using these models and investigated the shape and stability of AMPLATZER™ Duct Occluder/AMPLATZER™ Vascular Plug deployed in them. Results:Briefly, the time and cost required to create a hollow 3D model were about 3 days and 10,000 Japanese Yen, respectively. It was very easy to recognize the shape and stability of deployed devices in these models because the devices were clearly visible from outside. In fact, these models could be picked up by hands and observed from various angles, which was helpful in selecting optimum devices for closure. Conclusions:The transparent silicone hollow models are an excellent simulator of percutaneous closure of PDA, especially in adults. |
書誌情報 |
自治医科大学紀要
en : Jichi Medical University Journal
巻 41,
p. 21-28,
発行日 2019-03
|
出版者 |
|
|
出版者 |
自治医科大学 |
ISSN |
|
|
収録物識別子タイプ |
PISSN |
|
収録物識別子 |
1881-252X |